Overview of velashape
SHR stands for Super Hair Removal. Adopts Optimal Pulse Technology (OPT),it is the most advanced adoption of pulsed light for hair removal and skin rejuvenation.The working principle is heating the dermis to a target temperature gradually, the target temperature damages the hair follicles and prevents re-growing of them.
Treatment Theory of IPL SHR machine
A. Hair Removal:
The improved IPL with Flat-topped square wave emits varied wavelength, broad spectrum and intense pulse light;Flat-topped square wave means the energy is more stable and high; which permeate epidermis to the dermis of the skin;Utilizing the selective absorption elements, the light is absorbed by the melanin in the hair follicles. Through the optical & thermal effect generated and sourced from the light, it achieves the purpose of removing unexpected hair by rapid destroying of hair follicle tissues.
B.Skin rejuvenation:
The improved super IPL system emits varied wavelength, intense pulse, broad spectrum light;
It can permeate the cuticle to the dermal and take effect on the abnormal pigment and capillary vessel;
to break the abnormal pigment cells, close the abnormal blood vessels,stimulate the proliferation of collagen and improving the rearrangement of elastic fibers;
Finally achieving the purpose of pigment removal and skin rejuvenation, while keep normal skin and sweat gland in good state.
Function of IPL SHR machine
1. Skin rejuvenation: shrink pores, yellow and white, etc.
2. Freckle: epidermal spots, freckles, spotted spots, age spots, pigmentation, and sun damage and photoaging
3. Red blood: rosacea, facial flushing, facial red blood
4. Hemorrhoids: new acne and hemorrhoids
5. Hair removal: all kinds of body hair, mane, lip hair, hair, bikini, etc.
IPL SHR Machine Parameters
Type | IPL SHR |
IPL+ RF | Yes |
Certification | CE,ISO |
Place of Origin | Zhengzhou, China |
Brand Name | Mescien |
Model Number | IPL SHR machine |
Feature | Acne Treatment, Blood Vessels Removal, Hair Removal, Pigment Removal, Skin Rejuvenation, Skin Whitening |
Item name | Portable Elight Ipl Shr Mini Skin Whitening Machine |
Treatment Handle | Single Handle Tool(SR:530nmn-950nm or 640nm-950nm) |
Power | 1200W |
Spot Size | 10nm*50nm |
Energy Density | 1- 50 J/cm2, stepping 1J |
Cooling Method | Air+Water+Semiconductor |
Dimension | 69 * 49 * 58 cm |
Gross Weight | about 34kgs |
Xenon Lamp | UK First Light ( 300,000 effective shots) |
Keywords | Portable Elight Ipl Shr Mini Skin Whitening Machine |
Standard Operating Procedure (E-light )
Functions
Step l: Clean the customer 's treatment area.
If this is the first time for the customer do E light treatment, have a photosensitive test behind the customer's ear and wait for five minutes to check weather she is photosensitive skin, photosensitive skin people can not do E light treatment.
If the customer do hair removal treatment, cut the customer 's hair to about 2-3mm.
Step 2: Both the operator and the customer wear the safety glasses and put gel on the customer’s treatment area.
Step3:Choose proper therapy handle according to operation item,then touch the functions on the screen and click "GO". Keep the RF Energy to 18 and change IPL energy to the' start energy. generally about 2 to 4 J/CM2.
Touch the cooling beside the RF energy, the snowflake become blue means open, we suggest to light all five snowflake in summer and around 3 snowflake in winter also need to do according to customer's feeling.
Step4:Choose proper therapy handle according to operation item,then touch the functions on the screen and click "GO". Keep the RF Energy to 18 and change IPL energy to the' start energy. generally about 2 to 4 J/CM2.
Touch the cooling beside the RF energy, the snowflake become blue means open, we suggest to light all five snowflake in summer and around 3 snowflake in winter also need to do according to customer's feeling.
Step5: Clean gels on the treatment area. If you operate on whole face, suggest to use cool facial mask at once to help the face dissipate heat and also give water to the face. 1f you treat on other part, use ice bag or cold towel to cover the treatment area for 5 to 10 minutes.
Attention for E light treatment
1.Skin malignant tumors or precancerous lesions 2.pregnancy
3.sunburn skin recently
4.Photosensitivity diseases (lupus erythematosus, epilepsy) 5.Acute skin infections (bumps or frostbite)
6.Take photosensitive drugs recently (within one month)
After taking some medication, under the light stimulus, could cause human body allergic reactions, these drugs called photosensitivity drugs.
The photosensitivity of common drugs are: such as antibiotic tetracycline, doxycycline (doxycycline); Sulfa sulfadiazine; Quinolone class of norfloxacin, ofloxacin, ciprofloxacin, si pasha magnitude; Antifungal drugs such as griseofulvin. Oral medications D - 860; Sedatives of promethazine, chlorpromazine; Diuretic furosemide, hydrochlorothiazide, ammonia, benzene pteridine, etc.
7.Poor dry skin 8.Scar constitution
9.upper eyelid
10.Patients with diabetes, heart disease, and patients have unrealistic expectations.
11.Used effective products, the customer need to stop at least one month or more time, and do not use these products in the process of E light therapy. 12.Have implant parts.
13.Device with cardiac pacing, or built-in defibrillator.
If you want know more , please see this website : www.ms-laser.com or contact this whatsapp : +86 156 7023 3721